• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMEDA, INC. PURELY YOURS; ELECTRIC BREAST PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMEDA, INC. PURELY YOURS; ELECTRIC BREAST PUMP Back to Search Results
Model Number 24502082
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/12/2018
Event Type  malfunction  
Manufacturer Narrative
The returned product was tested per ameda engineering protocol to determine whether the allegation of leaking fluid could be confirmed.The returned product was assessed for indications of malfunction or thermal event.The returned product was assessed for functionality and met functional specifications.Dark grey substance, e.G battery acid, was observed inside the battery compartment.Internally, no signs of foreign substance other than battery fluid, burning, melting or charring were observed.The battery fluid crust was observed on returned batteries.
 
Event Description
Customer contacted ameda, inc.On (b)(6) 2018 to report clear/white battery fluid leaking from her purely yours breast pump base during a pumping session on (b)(6) 2018.Customer was using the ac adapter and had 6 aa batteries inside the battery compartment.Batteries were new out of the package.Customer reports this pump is out of warranty and used with her first baby 3 years prior.She states that when using the batteries she would have to turn her pump base upside down and tap on it to get it powering on.She reports no injury or burn in this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PURELY YOURS
Type of Device
ELECTRIC BREAST PUMP
Manufacturer (Section D)
AMEDA, INC.
485 half day rd.
suite 320
buffalo grove IL 60089
Manufacturer Contact
linda zager
485 half day rd.
suite 320
buffalo grove, IL 60089
8479642620
MDR Report Key7332194
MDR Text Key102351760
Report Number3009974348-2018-00311
Device Sequence Number1
Product Code HGX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973501
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Remedial Action Replace
Type of Report Initial
Report Date 02/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number24502082
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/12/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age30 YR
-
-