MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AESPIRE 7900 ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AESPIRE 7900 ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Detachment Of Device Component (1104)
Patient Problem No Patient Involvement (2645)
Event Date 02/13/2018
Event Type  Malfunction  
Manufacturer Narrative

A ge healthcare service representative performed a checkout of the equipment and confirmed the reported complaint. The caster base was replaced, and the unit was returned to service. No report of patient involvement. The initial reporter is located outside the u. S. , and therefore this information is not provided due to country privacy laws.

 
Event Description

The hospital reported the caster wheel became detached from the unit. There was no report of the unit tipping or falling. There is no report of injury.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAESPIRE 7900
Type of DeviceANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 
MDR Report Key7333676
MDR Text Key102205474
Report Number2112667-2018-00531
Device Sequence Number1
Product Code BSZ
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK050626
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL,USER FACI
Reporter Occupation BIOMEDICAL ENGINEER
Type of Report Initial
Report Date 03/13/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received03/13/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Was Device Available For Evaluation? Yes
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received02/13/2018
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured03/23/2010
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-