Additional manufacturer narrative: the device was not returned to edwards for evaluation.The clamp device is essential to occlude the aorta and provide the necessary cardiac isolation required to perform minimally invasive cardiac surgery procedures.If the balloon bursts during a procedure, the heart would fill and warm, the operative site may be obscured and the procedure may need to convert to an open procedure, the risk of injury is not remote.A definitive root cause could not be determined; however, it is likely that surgeon technique contributed to the vent as it was reported that the surgeon accidentally punctured the balloon during suture placement.As there has been no confirmed design defect, manufacturing issue, or inadequacy in the labeling, ifu, or training leading to the complaint, edwards will continue to review all reported events and perform trend analysis on a monthly basis.If action is required based on the determined control limits, appropriate investigation will be performed.No corrective or preventative actions are required at this time.
|
Edwards received notification that during procedure the balloon of an intraclude device was accidentally punctured.The issue occurred during placement of the sutures at the level of the anterior leaflet, during a mitral valve repair that was performed in mini thoracotomy.Another intraclude device was used in replacement.During replacement, the heart was fibrillating shortly and there was no blood loss.There was no reported complication for the patient.
|