Model Number S7
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 12/03/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Unique device identification (udi) is unavailable.Device manufacturing date is unavailable.A medtronic representative went to the site to test the equipment.It was reported that the computer for the navigation system was replaced.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.No parts have been received by the manufacturer for evaluation.
 
Event Description
A site representative reported that, while outside of a procedure, the navigation system would intermittently restart without prompt from the user.Clearing the patient archives and re-installation of application software did not restore functionality.There was no patient present when this issue was identified.No additional information was provided.
 
Manufacturer Narrative
Additional information: unique device identification (udi) and device manufacture date provided.
 
Manufacturer Narrative
The computer was returned to the manufacturer for analysis.Analysis found that the computer would not power on.When the computer was connected to a test power supply, the computer passes seatools, memtest86 and phd without errors.The computer was found to have a bad powers supply.Analysis found that the reported event was related to a electrical issue.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7336349
MDR Text Key102315768
Report Number1723170-2018-01119
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 12/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733858
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/21/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/19/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-