A field service engineer arrived at the customer site on 12/30/2015 to the address the reported event.The fse found an object inside the loader and removed a wood stick,cleaned and lubricated the sample loader and performed loader alignment verification.Several sample (x10) were run to verify the instrument, ran controls and ran 10 patient samples.Quality controls (qc) were acceptable.Chromatograms also passed.The instrument is working good now,the service is completed and no further action required by field service.The most probable cause of the reported event was due to a sample loader stuck due to a fallen object.(b)(4).This report is being submitted due to a retrospective review conducted under capa-(b)(4).
|