Catalog Number 72200616SR
Device Problem Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/26/2018
Event Type  malfunction  
Event Description
It was reported that the middle oscillating button was stuck, causing that the mdu continually run unintended. No significant delay or patient injury reported.
 
Manufacturer Narrative
The device was received for evaluation. There was no relationship found between the returned device and the reported incident. A visual inspection was performed on the product and no issues were observed. A functional evaluation did not reveal any problems. The oscillate button performed as expected and the unit passed testing. The complaint was not verified and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSVCE REPL MDU CNTRL PWRMX ELITE
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7340004
MDR Text Key102582775
Report Number1643264-2018-00225
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 04/19/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number72200616SR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/14/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-