DEPUY SYNTHES PRODUCTS LLC SMALL BATTERY DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED
|
Back to Search Results |
|
Catalog Number 532.010 |
Device Problem
Device Inoperable (1663)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|
Event Description
|
It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the trigger on the small battery drive device didn't activate the device anymore.It was reported that there was a 10 minute delay in the surgical procedure.It was reported that a spare device was available in order to complete the surgery with no further issues.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The actual device was returned for evaluation.Reliability engineering evaluated the device and determined that the bottom trigger came loose and was out of position, unknown residue in the electronic control unit and on triggers, motor makes an unusual noise while running, worn out coupling head and worn bearings and seals.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to improper cleaning and normal wear.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|