• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem Missing Value Reason (3192)
Event Date 05/16/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4) this report is being submitted due to a retrospective review conducted under capa (b)(4).Fse (field service engineer) was dispatched and on (b)(6) 2017 was the third time at customer site for same z1 issue.Very intermittant.Replaced possible bad asm board that could be causing issue.Placed in new sample needle.Ran 3 racks of patients with all good results.The most probable cause of the reported event was a faulty asm board.
 
Event Description
On (b)(6) 2017, the customer reported z1 axis error with their hlc-723g8 analyzer.The customer is unable to run hba1c patient samples.A field service engineer (fse) was dispatched to address the reported event, which resulted in delayed reporting hba1c patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION (MANUFACTURER)
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key7341537
MDR Text Key103149965
Report Number8031673-2018-02532
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/15/2018
Distributor Facility Aware Date05/16/2017
Device Age2 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer03/15/2018
Date Manufacturer Received05/16/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-