Catalog Number 03.114.001
Device Problem Fitting Problem (2183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/13/2018
Event Type  malfunction  
Manufacturer Narrative
Device used in a veterinary case - no patient information will be reported.Device is an instrument and is not implanted/explanted.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.A device history record (dhr) review was performed for: part number: 03.114.001 , synthes lot number: h288454 , supplier lot number: (b)(4).Release to warehouse date: 27-feb-2017 , supplier: (b)(4), manufacturing site: (b)(4).No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes on an event in (b)(6) as follows: this report is for a veterinary case.There was no human patient involvement.It was reported, that the drill sleeve doesn't fix the plates and screws, therefore, they don't assembly properly.This complaint involves 1 part.This report is 1 of 1 for (b)(4).
 
Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.3008812560-08/04/2017-001-c device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: part: 03.114.001; synthes lot: h288454; supplier lot: h288454; release to warehouse date: february 27, 2017, supplier: flextronics america, llc; manufacturing site: monument.No nonconformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.Appropriate action has been initiated on june 6, 2017 for lcp drill sleeve 03.114.001 does not engage with plate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Concomitant devices: plate (part: unknown, lot: unknown, quantity: 1), screws (part: unknown, lot: unknown, quantity: unknown).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.1MM LCP THREADED DRILL GUIDE FOR 1.5MM LCP PLATES
Type of Device
GUIDE
Manufacturer (Section D)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key7342106
MDR Text Key102878658
Report Number8030965-2018-52099
Device Sequence Number1
Product Code FXZ
UDI-Device Identifier07611819393967
UDI-Public(01)07611819393967(10)H288454
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 02/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.114.001
Device Lot NumberH288454
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2019
Date Manufacturer Received02/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-0669-2018
Patient Sequence Number1
-
-