MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008T HEMODIALYSIS SYS., WITH CDX DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008T HEMODIALYSIS SYS., WITH CDX DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 190713
Device Problems Thermal Decomposition of Device (1071); Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/13/2018
Event Type  malfunction  
Manufacturer Narrative
A supplemental medwatch report will be submitted upon completion of the investigation.
 
Event Description
A user facility biomedical engineer (biomed tech) reported a 2008t machine powered off during patient treatment. Additionally, the machine blew fuses upon power up and the power supply fan did not come on. The biomed tech reported the insulation covering of the wire to the bridge rectifier was discolored and burnt. The part was discarded and is not available to be sent back for evaluation. No photos were taken of the burnt component. The bridge rectifier was replaced to resolve the reported issues. The patient on the machine during the reported event was able to complete treatment on the same machine and did not experience any adverse effects or require medical intervention as a result of this incident. The machine was returned to service.
 
Event Description
A user facility biomedical engineer (biomed tech) reported a 2008t machine powered off during patient treatment. Additionally, the machine blew fuses upon power up and the power supply fan did not come on. The biomed tech reported the insulation covering of the wire to the bridge rectifier was discolored and burnt. The part was discarded and is not available to be sent back for evaluation. No photos were taken of the burnt component. The bridge rectifier was replaced to resolve the reported issues. The patient on the machine during the reported event was able to complete treatment on the same machine and did not experience any adverse effects or require medical intervention as a result of this incident. The machine was returned to service.
 
Manufacturer Narrative
Plant investigation: the device was not returned to the manufacturer for physical evaluation and the failure mode cannot be confirmed. However, an investigation of the device manufacturing records was conducted by the manufacturer. There were no deviations or non-conformances during the manufacturing process. All device history records (dhr) are reviewed and released according to the dhr review checklist and release procedure. A device is not released if it does not meet requirements or is nonconforming. In addition, the device record review confirmed the labeling, material, and process controls were within specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name2008T HEMODIALYSIS SYS., WITH CDX
Type of DeviceDIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
MDR Report Key7345155
MDR Text Key102627605
Report Number2937457-2018-00793
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
PMA/PMN Number
K093902
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 03/22/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/16/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number190713
Was Device Available for Evaluation? No
Was the Report Sent to FDA?
Device AgeMO
Event Location No Information
Date Manufacturer Received03/21/2018
Is This a Reprocessed and Reused Single-Use Device? No

-
-