DEPUY SYNTHES PRODUCTS LLC COLIBRI II HANDPIECE INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.101
Device Problems Unintended System Motion (1430); Device Stops Intermittently (1599); Fail-Safe Problem (2936)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis. (b)(6). This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment. During repair, it was determined that the motor of the device had seized, jammed, and was moving heavily. It was noted that the handpiece had no response upon triggering. Isolated and found motor malfunction. It was further determined that the device failed pretest for check the triggers and electric control unit (ecu) function, check switching function, check the oscillation angle, check the off/oscillation/on switch mode function. Therefore, the reported condition was confirmed. The assignable root cause was determined to be due to wear from normal use and servicing. If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Event Description
It was reported from (b)(6) that the small battery drive device was not completely functioning. It was noted that sometimes it would work and sometimes it would not work. It was not reported if the device was used in surgery, or if there was patient involvement. It was not reported if there were any delays in a surgical procedure or if a spare device was available. It was not reported if there were any injuries, medical intervention or prolonged hospitalization. The exact date of the event was unknown. All available information has been disclosed. If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCOLIBRI II HANDPIECE
Type of DeviceINSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ 4436
Manufacturer (Section G)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ 4436
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key7348208
MDR Text Key102758204
Report Number8030965-2018-52177
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeTH
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial
Report Date 02/21/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number532.101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2018
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/21/2018
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured11/05/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-