• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY FRANCE SAS 3003895575 CORAIL AMT COLLAR SIZE 13; CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS 3003895575 CORAIL AMT COLLAR SIZE 13; CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM Back to Search Results
Catalog Number 3L92503
Device Problem Migration or Expulsion of Device (1395)
Patient Problems Bone Fracture(s) (1870); Pain (1994)
Event Date 02/23/2018
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Der states that the right painful hip, corail stem subsided 3-4 mm, medial calcar crack, cabled calcar above and below lesser troch.New longer 15.5 head was used.Doi: (b)(6) 2018; (b)(6) 2018; right hip.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL AMT COLLAR SIZE 13
Type of Device
CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY FRANCE SAS 3003895575
7 allée irène joliot curie
bp 256
saint priest cedex 69801
FR  69801
Manufacturer (Section G)
DEPUY INTERNATIONAL LTD. 8010379
st. anthony's rd
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46582
5743725905
MDR Report Key7349350
MDR Text Key102763850
Report Number1818910-2018-55451
Device Sequence Number1
Product Code KWL
UDI-Device Identifier10603295168690
UDI-Public10603295168690
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123991
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 02/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number3L92503
Device Lot Number5297668
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/04/2018
Date Device Manufactured07/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age68 YR
Patient Weight113
-
-