• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 9500-45
Device Problem Component Missing (2306)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/18/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4). The patient experienced adverse events on different days with the same device. The following reports are being submitted. (b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2018 that on (b)(6) 2018, a missing sensor wire occurred. The sensor was inserted into the arm on (b)(6) 2018. No additional event or patient information is available. No product was provided for evaluation. The complaint confirmation was unable to be determined. A root cause could not be determined. Labeling indicates: do not insert the sensor component of the g5 mobile system in a site other than the belly/abdomen (ages 2 years and older) or the upper buttocks (ages 2 to 17 years). The placement and insertion of the sensor component of the g5 mobile system is not approved for other sites. If placed in other areas, the g5 mobile system may not function properly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of DeviceCONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key7351147
MDR Text Key102864438
Report Number3004753838-2018-27683
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Reporter Occupation
Type of Report Initial
Report Date 02/19/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date06/28/2018
Device Model Number9500-45
Device Catalogue NumberSTS-GF-004
Device Lot Number5226906
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received02/19/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-