Boston scientific received information that a lead safety switch (lss) was declared for both the right ventricular (rv) lead and left ventricular (lv) lead.
The pacing impedance for both leads was greater than 2000 ohms.
The impedances then went back to a normal range.
It was noted there was a lot of noise present after the lss triggered.
The noise could be reproduced with isometrics.
The noise was oversensed, which resulted in asystole of greater than two seconds.
The physician planned to program the device back to a bipolar configuration because the patient is pacer dependent.
No further actions were taken.
The root cause for the clinical observations was not determined.
Additional information was received that further review of the patient¿s data found noise episodes from four months earlier.
The noise presented on the right atrial (ra) and rv channels and was oversensed.
Boston scientific technical services (ts) reviewed the electrograms (egms) and stated the noise was from the cardiac resynchronization therapy pacemaker (crt-p) oversensing the minute ventilation signal.
Ts discussed how to program the rate response trend and minute ventilation features off.
It was also noted that another manufacturer¿s ra lead impedance measurement was out of range two years earlier.
However, currently the ra lead impedance measurements were within range.
The patient was scheduled to have the minute ventilation feature programmed off.
The crt-p and all leads remain in service and no adverse patient effects were reported.
|