Model Number EX1150
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/05/2018
Event Type  malfunction  
Manufacturer Narrative
The results of the analysis confirmed the reported event of failure to start up.Inspection of the ac power adapter revealed burn marks on the main circuit board and the inner casing of the ac power adapter consistent with damage due to the high current flow through the ac wall power outlet.
 
Event Description
This report is to advise of an event observed during analysis.Burn marks to the main pcb were revealed during analysis.There were no reported adverse patient consequences related to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN@HOME, RF TELEMETRY BASIC
Type of Device
PACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7353807
MDR Text Key102945277
Report Number2017865-2018-03360
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
Reporter Country CodeEZ
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Patient
Type of Report Initial
Report Date 03/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/20/2018
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberEX1150
Device Catalogue NumberEX1150
Other Device ID Number05414734504799
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received02/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/26/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-