Model Number U228
Device Problems Over-Sensing (1438); Pacing Problem (1439); Ambient Noise Problem (2877)
Patient Problem Dizziness (2194)
Event Date 02/23/2018
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete. This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this patient reported having dizzy spells. Upon interrogation, noise episodes were observed with were consistent with minute ventilation oversensing resulting in pacing inhibition. As a result, the minute ventilation sensor was programmed off. No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameACCOLADE
Type of DeviceIMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7358064
MDR Text Key103052950
Report Number2124215-2018-04268
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeGB
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial
Report Date 02/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date03/27/2019
Device Model NumberU228
Other Device ID NumberVISIONIST X4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Event Location No Information
Date Manufacturer Received02/23/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/27/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 03/21/2018 Patient Sequence Number: 1
Treatment
7741; 7742; U228
-
-