It was reported that during a cryoablation procedure, air was observed in the sideport of the sheath while the sheath was placed in the left atrium.The physician tried to aspirate but only air from the sheath valve was removed.An x ray was performed and the physician observed a massive air embolism in the patient's left atrium, left ventricle and aorta.The guide wire was removed and then it became possible to aspirate the air.It was noted that the patient experienced hemiparesis of the arm, very low la pressure and started to breath very deep.The physician reported that the valve of the sheath with the wire did not prevent air ingress.The procedure was aborted and the patent was hospitalized.No further patient complications have been reported as a result of this event.
|
Event summary: visual inspection of flexcath sheath 4fc12 / 22147-097, results showed the device was intact with no apparent issues.Functional testing of the sheath confirmed the hemostatic valve was leaking.Air bubbles were observed through the valve.It was suspected that the valve disk was torn.This is a clinical issue encountered during the procedure.In conclusion, the reported (embolism, hypotension, shortness of breath, paralysis or paresis) issue cannot be confirmed through testing.The reported (air ingress during aspiration, hemostatic valve issue) issue was confirmed through testing.The reported (sideport aspiration issue) issue was not confirmed through testing.This is a clinical issue encountered during the procedure.The sheath failed the returned product inspection due to a leaking hemostatic valve.If information is provided in the future, a supplemental report will be issued.
|