GUIDANT CRM CLONMEL IRELAND ACCOLADE IMPLANTABLE PULSE GENERATOR Back to Search Results
Model Number U128
Device Problems High impedance (1291); Over-Sensing (1438); Device Sensing Problem (2917)
Patient Problems Atrial Tachycardia (1731); No Consequences Or Impact To Patient (2199)
Event Date 10/14/2017
Event Type  malfunction  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete. This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy pacemaker (crt-p) triggered an alert for high out of range pace impedances on the right ventricular (rv) lead. At implant, the impedances were 398 ohms and were now greater than 3000 ohms. Further review noted that there had also been oversensing of the minute ventilation (mv) signal prior to the out of range impedance alert. The mv oversensing resulted in pacing inhibition, but there was no asystole due to the patient's intrinsic rhythm. In addition, when the patient was in an atrial arrythmia, the left ventricular (lv) lead was picking up the signal, but not oversensing it. At this time, the products remain in service and the patient is being monitored. The rv lead is a competitor's product. There have been no adverse patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameACCOLADE
Type of DeviceIMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7362308
MDR Text Key103277154
Report Number2124215-2018-05302
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial
Report Date 02/21/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/22/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date06/30/2018
Device Model NumberU128
Other Device ID NumberVALITUDE X4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Event Location No Information
Date Manufacturer Received02/21/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/30/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-