Model Number S7 |
Device Problems
No Display/Image (1183); Device Stops Intermittently (1599)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/28/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient weight not available from the site.
A medtronic representative went to the site to test the equipment.
The representative was unable to replicate the reported issue.
The hardware, software, and instruments passed the system checkout.
The system was found to be fully functional.
|
|
Event Description
|
A medtronic representative reported that, while in a transsphenoidal resection procedure, the monitor of the navigation system lost display functionality.
It was reported that re-plugging the monitor in temporarily restored functionality.
There was no reported delay to the procedure due to this issue.
There was no reported impact on patient outcome.
No additional information was provided.
|
|
Manufacturer Narrative
|
Additional information: patient weight provided.
Correction: initial reporter if information is provided in the future, a supplemental report will be issued.
|
|