DATASCOPE MAHWAH CS100; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problems Device Alarm System (1012); Device Stops Intermittently (1599)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/18/2018
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was not required to be reviewed per company standard operating procedure since the device manufacture date is greater than one year from the event date.A field service engineer of the getinge authorized distributor evaluated the iabp and no problem was found.The iabp worked normally, no parts were replaced and all functional and safety tests were passed to factory specifications.The iabp was cleared for clinical use.Full name of the event site: (b)(6) hospital (b)(6).(b)(6).
 
Event Description
It was reported that during use, the cs100 intra-aortic balloon pump (iabp) stopped pumping without any human intervention and without sounding an alarm.The customer claimed that the alarm sound came after the iabp stopped pumping for almost 10 minutes.Also the iabp was unable to initiate assist by pressing the start key, but when the doctor found the problem, the iabp worked normally after the doctor restarted the iabp.Video depiction of the event was provided by the customer process with a request for explanation as to why the iabp stopped assisting and why the alarm was delayed during the event.No adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key7366177
MDR Text Key103550681
Report Number2249723-2018-00477
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial
Report Date 03/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0998-00-3013-XX
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Device AgeYR
Date Manufacturer Received02/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age89 YR
Patient Weight50
-
-