PURELY YOURS; ELECTRIC BREAST PUMP Back to Search Results
Model Number 24502082
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/26/2018
Event Type  malfunction  
Manufacturer Narrative
The returned product was tested per ameda engineering protocol to determine whether the allegation of leaking fluid could be confirmed.The returned product was assessed for indications of malfunction or thermal event.The returned product was assessed for functionality and met functional specifications.Dark grey substance, e.G battery acid, was observed inside the battery compartment.Internally, no signs of foreign substance, burning, or melting were observed.The returned batteries were observably damaged.The battery fluid crust was observed on returned batteries.Additional testing confirmed that batteries may leak when the upper middle battery is placed incorrectly, with the positive and negative end reversed.
 
Event Description
Customer contacted ameda, inc.On (b)(6) 2018 to report an incident today while using her purely yours breast pump on battery power.She only uses batteries while working one day per week.Customer states hearing a loud pop inside the battery compartment followed by leakage of milky, black tinged fluid inside the battery compartment.Fluid stayed contained inside the compartment.Customer denied any burn or injury in this event.A replacement purely yours breast pump was overnight shipped to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PURELY YOURS
Type of Device
ELECTRIC BREAST PUMP
Manufacturer Contact
linda zager
485 half day rd.
suite 320
buffalo grove, IL 60089
8479642620
MDR Report Key7367745
MDR Text Key103691956
Report Number3009974348-2018-00313
Device Sequence Number1
Product Code HGX
UDI-Device Identifier00810725021477
UDI-Public(01)00810725021477(21)
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973501
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Remedial Action Replace
Type of Report Initial
Report Date 02/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number24502082
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/26/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/18/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age35 YR
-
-