Model Number FUSION |
Device Problems
Device Stops Intermittently (1599); Communication or Transmission Problem (2896)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 03/05/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.
No parts have been received by the manufacturer for evaluation.
|
|
Event Description
|
A medtronic representative reported that, while outside of a procedure, the emitter would intermittently experience difficulties tracking instruments when there was no metal interference.
There was no patient present when this issue was identified.
No additional information was provided.
|
|
Manufacturer Narrative
|
The field generator was returned to the manufacturer for analysis.
The field generator was connected to a known good axiem test system for a twenty four hour burn-in test.
The system remained in green status throughout all testing.
The reported problem could not be duplicated.
|
|
Manufacturer Narrative
|
A medtronic representative went to the site to test the equipment.
Testing revealed that a field generator was replaced to resolve the issue.
The system then passed the system checkout and was found to be fully functional.
|
|