Model Number 8637-20
Device Problems Failure To Service (1563); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/12/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a health care provider (hcp) via a manufacturer representative regarding a patient with an implantable drug infusion pump indicated for spinal pain.The pump contained an unknown brand of morphine with an unknown concentration and dose [4 mg/ml] at a dose of 0.65 mg/day.It was reported an empty pump alarm was occurring.The patient missed a refill.The representative had access to a physician programmer.The representative stated the hcp did not want the patient to be getting any of the old drug, so they wanted to do a full system content removal.The new morphine concentration and dose were [1 mg/ml] and 0.1 mg/day.The event was reported to have occurred ¿two weeks ago¿ relative to the day of report ((b)(6) 2018).No patient symptoms/complications were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received from the hcp via the representative reported the issue was resolved as the pump was refilled and rep rogrammed to the new dose and concentration.The patient¿s weight at the time of the event was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7373799
MDR Text Key103647330
Report Number3004209178-2018-05816
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169508149
UDI-Public00643169508149
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/14/2018
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/27/2018
Date Device Manufactured01/18/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age58 YR
-
-