We originally reported 2 burs on mdr 1045254-2016-00374.A supplemental report for the product analysis to 1045254-2017-00342 was provided regarding the first bur, information on the other bur is being submitted in this report.This is 1 of 1 subsequent reports to be filed.Product evaluation: the hi speed diamond 70deg bur, part number 1883672hs, from lot number 0209955643, was received for analysis.There was a residue consistent with biological contaminants on the devices.Biological contaminants were found to be compacted in the diamond grit which likely resulted in the customer applying excess pressure to the bur in an attempt to maintain performance.When viewed under magnification there was evidence of aggressive use; gouging of the outside diameter of the inner shaft just proximal to the tip; and corresponding damage to the outer tube support area (including a thinning of the wall).Visually, the inner assembly tip was stretched out of the outer tube support area which would have resulted in the reported event.The spiral wrap was twisted in on itself in a clockwise direction.The location of the damage is consistent with excessive torsional load between the gouged area of the inner shaft and the spiral wrap proximal to the damage.If information is provided in the future, a supplemental report will be issued.
|