MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9735665
Device Problems Device Stops Intermittently (1599); Communication or Transmission Problem (2896); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/07/2018
Event Type  malfunction  
Manufacturer Narrative
No parts have been received by the manufacturer for evaluation.
 
Event Description
A manufacturer representative reported that, while in a spinal fusion, the navigation system was unable to establish communication with the medtronic imaging system.It was reported that a second ethernet cable was used without resolution.Restarting the imaging system restored functionality.There was no reported impact on patient outcome.No additional information was provided.
 
Manufacturer Narrative
Additional information: additional information received reported there was a delay of 10 minutes due to the reported issue.
 
Manufacturer Narrative
Additional information: software logs show that imaging system never established communication with navigation system.A restart of imaging system computer resolved issue.Software is functioning as designed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 PREMIUM SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
heather davis
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7377727
MDR Text Key103703267
Report Number1723170-2018-01339
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169722187
UDI-Public00643169722187
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date 05/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9735665
Is the Reporter a Health Professional? No
Date Manufacturer Received04/19/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/28/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age36 YR
Patient Weight95
-
-