Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Investigation - evaluation.Visual inspection and functional testing of the returned device was conducted during the investigation.A review of complaint history, the device history record, instructions for use, quality control data, and specifications was also performed.One device was returned for investigation.The device was returned trapped in the shipping tray.The device was returned with the handle in the open position.The basket formation is in the open position.The collet knob is tight and secure.The male luer lock adaptor (mlla) is loose.The polyethylene terephthalate tubing (pett) measures 2.5 cm in length.A visual examination noted kinks in the basket sheath at the following locations: 18 cm from the distal tip, 44.2 cm from the distal tip, and 84 cm from the distal tip.A functional test was performed and determined the handle does not actuate the basket formation.With the handle in the closed position, the basket assembly is accordion starting at 5 cm from the distal tip of the basket sheath and extends to 12 cm from the distal tip.The support sheath and the basket sheath are still adhered.The handle was disassembled and the basket formation could not be manually actuated.The device history record was reviewed and found there were no non-conformances related to the failure.A review of complaint history revealed this complaint is the only one associated with lot number 7002862.The instructions for use (ifu) provides the following instructions in the precaution section: enclose the device in the sheath before removing from the tray/holder.Do not use excessive force to manipulate this device.Damage to the device may occur.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was visually/functionally inspected by quality control and no related gaps in production or processing controls were noted.The complaint device was found to have a non-functioning basket due to sheath damage.Devices are inspected for damage and functionality prior to packaging.The observed damage likely occurred during handling/use of the device.Per the quality engineering risk assessment, no further action is warranted.Cook medical has notified the appropriate personnel and will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.H3 other text : blank fields on this form indicate the information is unknown, unchanged, or unavailable.Investigation - evaluation.Visual inspection and functional testing of the returned device was conducted during the investigation.A review of complaint history, the device history record, instructions for use, quality control data, and specifications was also performed.One device was returned for investigation.The device was returned trapped in the shipping tray.The device was returned with the handle in the open position.The basket formation is in the open position.The collet knob is tight and secure.The male luer lock adaptor (mlla) is loose.The polyethylene terephthalate tubing (pett) measures 2.5 cm in length.A visual examination noted kinks in the basket sheath at the following locations: 18 cm from the distal tip, 44.2 cm from the distal tip, and 84 cm from the distal tip.A functional test was performed and determined the handle does not actuate the basket formation.With the handle in the closed position, the basket assembly is accordion starting at 5 cm from the distal tip of the basket sheath and extends to 12 cm from the distal tip.The support sheath and the basket sheath are still adhered.The handle was disassembled and the basket formation could not be manually actuated.The device history record was reviewed and found there were no non-conformances related to the failure.A review of complaint history revealed this complaint is the only one associated with lot number 7002862.The instructions for use (ifu) provides the following instructions in the precaution section: enclose the device in the sheath before removing from the tray/holder.Do not use excessive force to manipulate this device.Damage to the device may occur.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was visually/functionally inspected by quality control and no related gaps in production or processing controls were noted.The complaint device was found to have a non-functioning basket due to sheath damage.Devices are inspected for damage and functionality prior to packaging.The observed damage likely occurred during handling/use of the device.Per the quality engineering risk assessment, no further action is warranted.Cook medical has notified the appropriate personnel and will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|