Catalog Number 05.001.000
Device Problems Device Operates Differently Than Expected (2913); Device Operational Issue (2914)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/08/2018
Event Type  malfunction  
Manufacturer Narrative
The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that during a shoulder latarjet surgical procedure, it was observed that the blue power light of the console device came on but would not respond to anything.There were no delays to the surgical procedure as a spare device was available to complete the surgery.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the reported condition was confirmed.It was determined that the housing was cracked and the back of the housing was broken.It was observed that the ¿sed¿ upgrade was missing.It was further determined that the device failed pretest for general condition, version (upgrade) check and check function of sed has been changed.The assignable root cause was determined to be due to improper handling, which is user error, misuse, and / or abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STANDARD CONSOLE WITH IRRIGATION FOR EPD
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key7378617
MDR Text Key103729135
Report Number8030965-2018-52521
Device Sequence Number1
Product Code HWE
UDI-Device Identifier7611819167681
UDI-Public(01)7611819167681(11)091016(10)1197
Combination Product (y/n)N
PMA/PMN Number
K093361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 03/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.000
Device Lot Number1197
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2018
Date Manufacturer Received04/19/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-