ZIMMER BIOMET, INC. NEXGEN POROUS CR-FLEX FEMORAL; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problems Fracture (1260); Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Information (3190)
Event Date 02/26/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint device has been returned, but the device investigation has not yet been completed.Once the evaluation is completed, a supplemental medwatch will be submitted.
 
Event Description
It was reported that the implant packing was found fractured while opening sterile container.No adverse events were reported as a result of this malfunction.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Inspection of the returned device revealed some damage to the box and the inner and outer cavities were cracked.Also, the box was opened on the wrong side as evidenced by the two-side spanning label that is torn at the corner of the two sides it spans.Device history record (dhr) was reviewed and no discrepancies were found relevant to the reported event.Investigation results concluded the most likely root cause of the event is shipping/transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEXGEN POROUS CR-FLEX FEMORAL
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7379729
MDR Text Key103799975
Report Number0001822565-2018-01831
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK173057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2024
Device Model NumberN/A
Device Catalogue Number00595201705
Device Lot Number62575645
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/17/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/17/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-