Model Number UNK744
Device Problems Migration or Expulsion of Device (1395); Device Damaged by Another Device (2915)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/05/2018
Event Type  Injury  
Manufacturer Narrative
Device is a combination product.(b)(4).Device evaluated by mfr: the device was not returned therefore analysis of complaint device could not be performed.The investigation conclusion is caused by other device as another device/drug/subsequent procedure caused the complaint event.(b)(4).
 
Event Description
Same case as mdr id: 2134265-2018-02243 it was reported that stent migration occurred.The 90% stenosed target lesion was located in the mildly tortuous and moderately calcified proximal to mid left anterior descending artery.Following pre-dilation with a 2.0x15 emerge balloon catheter, a synergy¿ drug-eluting stent was implanted for treatment.An opticross¿ imaging catheter was then inserted for intravascular ultrasound; however difficulties pulling back the imaging catheter were encountered.The stent strut was caught with the imaging catheter and the stent migrated.The procedure was completed with another of the same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY¿
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7381968
MDR Text Key103857015
Report Number2134265-2018-02675
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeKR
PMA/PMN Number
SIMILAR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 03/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUNK744
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/05/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age66 YR
-
-