Model Number 728306 |
Device Problem
No Audible Alarm (1019)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.Internal cross reference: complaint pr# (b)(4).
|
|
Event Description
|
This complaint has been evaluated based on the information provided; there is no allegation of death or serious injury.The customer reported they are not able to hear the patients from the scan room.The philips field service engineer (fse) confirmed the auto-voice commands and the operator can be heard by the patient, but the operator can not hear the patient.If the operator is unable to hear the patient due to a failed microphone, there is potential for injury to the patient.This issue has been determined to be a reportable event.
|
|
Manufacturer Narrative
|
A philips field service engineer (fse) found that a microphone wire was loose on one of the four microphones in the gantry.When the fse reconnected the wire, a buzzing noise occurred.The fse confirmed that there was no patient injury; this event occurred during corrective maintenance and there was no patient involvement.The fse determined that the buzzing noise could not be isolated to any one microphone in particular, and was only present if all four microphones were plugged in.Unplugging any one of the four microphones would cause the buzzing noise to subside.The fse replaced all four microphones located on the gantry, the gantry audio board, along with all associated cables from the rear cone to the gantry to resolve the issue.The system is operational and in clinical use.This event has been determined not to be a reportable event.
|
|
Search Alerts/Recalls
|
|
|