U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE; PROSTHESIS, MITRAL VALVE, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE; PROSTHESIS, MITRAL VALVE, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9600TFX23
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Stroke/CVA (1770)
Event Date 12/20/2017
Event Type  Injury  
Manufacturer Narrative
Investigation is ongoing.
 
Event Description
As reported by the edwards affiliate in (b)(6), during a transaxillary tavr procedure in the mitral position, post deployment of a 23mm sapien 3 valve, the patient had a stroke.The patient was noted to have expired post procedure due to ¿toxicity caused by refusal of dialysis.¿.
 
Manufacturer Narrative
The patient¿s medical history was requested, however, the information was not forthcoming.The valve remains implanted in the patient and was therefore not returned to edwards for evaluation.Per the instructions for use (ifu), permanent or transient neurological events including stroke are potential adverse events associated with the tavr procedure and the use of the edwards thv devices.According to literature review, and as documented in a clinical technical summary, written by edwards lifesciences, stroke is recognized in the literature as a well-known complication in a small number of patients undergoing tavr.Risk factors correlating with a number of patient co-morbidities have been identified.Although in many cases the root cause of the event is unable to be determined, strokes during tavr are undoubtedly multifactorial, the dominant etiology likely being intra-procedure embolic events.A transcranial doppler study during tavr demonstrated that the majority of procedural embolic events occurred during balloon valvuloplasty, manipulation of catheters across the aortic valve, and valve implantation.An analysis in patients undergoing valve surgery revealed four baseline characteristics and two procedural events that were associated with early post-procedure stroke: female sex, ef < 30%, diabetes, age older than 70 years, bypass procedure time> 120 min, and calcification of the ascending aorta.Predictors of late stroke have included female sex, age older than 75 years, atrial fibrillation, and a history of or current smoking.There were no important differences in the frequency of late strokes between tavr and avr patients.After tavr, there appears to be a more significant proportion of early strokes occurring < 24 h post-procedure, but tavr patients with multiple co morbidities are probably at higher risk of both early and late strokes.In this case, there was no allegation or indication a device malfunction contributed to this adverse event.The cause of the stroke is unknown, however, may be due to patient factors (not provided) or the mechanisms described above.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE
Type of Device
PROSTHESIS, MITRAL VALVE, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92618
MDR Report Key7385073
MDR Text Key103947598
Report Number2015691-2018-01180
Device Sequence Number1
Product Code NPU
Combination Product (y/n)N
PMA/PMN Number
P140031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date04/24/2019
Device Model Number9600TFX23
Was Device Available for Evaluation? No
Date Manufacturer Received04/13/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-