Model Number 72200616S
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/01/2018
Event Type  malfunction  
Event Description
It was reported that the shaver not functioning. No back up was available.
 
Manufacturer Narrative
The device was received for evaluation. A visual inspection was performed on the exterior of product and no physical damage was observed. There was a relationship found between the returned device and the reported incident. Mdu failed functional tests with motor stall error and overheating when power cord was bent near strain relief. Power cord assembly grew warm during testing. After troubleshooting, the cause of motor stall and overheating was observed to be a defective power cord assembly pn: (b)(4). A visual inspection was performed on the power cords external covering and no physical damage was observed. It was determined that the power cord has shorted or open internal wiring. The complaint investigation has concluded that this unit has succumbed to physical damage to power cord. Factors which can contribute to a shorted power cord include rolling a heavy cart over the cord or excessive bending of the cord. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSVC REPL MDU HAND CNTRL PWRMX
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7392101
MDR Text Key104414115
Report Number1643264-2018-00265
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/26/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/03/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number72200616S
Device Catalogue Number72200616S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/23/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-