Model Number S7 |
Device Problems
No Display/Image (1183); Device Stops Intermittently (1599)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/12/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information not provided due to (b)(6) patient privacy regulations.
No procode, common device name, unique device identification (udi) and/or 510k provided as this device is not released for distribution in the united states.
A medtronic representative went to the site to test the equipment.
It was reported that the computer of the navigation system was replaced to restore functionality.
|
|
Event Description
|
A medtronic representative reported that, while in a procedure, the navigation system monitor blacked out without prompt from the user.
It was reported that restarting the navigation system temporarily restored functionality.
The monitor would then lose functionality later in the procedure.
The reported issue occurred four times in the procedure and restarting the navigation system would be performed until the procedure was completed.
There was no reported impact on patient outcome.
No additional information was provided.
|
|
Manufacturer Narrative
|
The computer was returned to the manufacturer for evaluation and after functional testing and visual/physical examination, no fault was found.
The issue could not be reproduced.
|
|