Model Number S7
Device Problem No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Date 03/12/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative went to site to test the equipment.Representative was unable to replicate the reported issue.Medtronic representative reported that the system was left overnight and site confirmed that the no other issue was noted.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.No parts were replaced.No parts have been received by the manufacturer for evaluation.
 
Event Description
A medtronic representative reported that while outside of procedure, the monitor of the navigation system displayed black and could not be turned on.There was no patient present at the time of the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
prashanth gali
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7395303
MDR Text Key104274071
Report Number1723170-2018-01421
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 04/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/04/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733856
Is the Reporter a Health Professional? No
Date Manufacturer Received03/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/27/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-