This device was being used off label to place a stent.
There is a specific warning in the instructions for use that states - "the catheter is not intended for use with stents.
" this catheter is only approved for percutaneous transluminal valvuloplasty.
Both the catheter and the stent were being used off label for unapproved indications.
It states in the report from the physician that he over inflated the balloon to 3.
0 atm.
The labeled rbp for this size is 2.
0 atm.
A review of the complaint device shows that a pinhole is located in the balloon.
The pin hole was most likely caused by the crimping of the stent on the balloon.
This is what caused the balloon to not inflate, which in turn caused the stent not to deploy properly.
Two comparative catheters - same model and balloon size were tested for rated burst pressure.
One of the comparative catheters was taken to 3.
5 atm before burst, and the other comparative catheter was taken to 4.
0 atm before burst.
The off label use, failure to follow instructions, and user error caused this adverse event and product malfunction.
|