• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL - DIABETES CARE BD ULTRA-FINE II INSULIN SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL - DIABETES CARE BD ULTRA-FINE II INSULIN SYRINGE Back to Search Results
Catalog Number 326702
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/12/2018
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Investigation results: summary: flks received an empty envelope that was torn open on two sides with nothing inside the envelope.It appears that anything that may have been sent in by the customer may have fallen out of the envelope during shipping.No samples were returned therefore the complaint could not be confirmed and the root cause is undetermined.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.As per manufacturing, a review of the device history record was completed for batch # 7079914.All inspections and challenges were performed per the applicable operations qc specifications.Based on the above, no additional investigation and no capa is required at this time.Conclusion: based on the samples / photo(s) received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Root cause: root cause cannot be determined at this time as the issue is unconfirmed as no samples or photos were returned.
 
Event Description
It was reported by a consumer that difficult plunger movement was found while using a bd ultra-fine ii insulin syringe.There was no report of exposure, injury or medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ULTRA-FINE II INSULIN SYRINGE
Type of Device
INSULIN SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key7400624
MDR Text Key104758981
Report Number1920898-2018-00188
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K955235
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date03/20/2017
Device Catalogue Number326702
Device Lot Number7079914
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-