Model Number 2432-0007
Device Problem Fluid/Blood Leak (1250)
Patient Problems Fever (1858); Unspecified Infection (1930)
Event Type  Injury  
Manufacturer Narrative
The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
 
Event Description
The customer reported that they found 0.9% ns leaking at the filter of the tubing.The patient spiked a fever and the blood culture results were positive for infection.There was no report of lasting harm.
 
Manufacturer Narrative
The customer¿s report of leaking at the filter was confirmed.Visual inspection under magnification showed no holes/tears in the filter membrane or damage to the filter housing.Functional and pressure testing confirmed leaking from the filter vent closest to the output port.The root cause of the filter vent leak was identified as the fluid resistance of the filter becoming roughly equivalent to the fluid resistance of the air vent after becoming clogged from prolonged exposure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS PUMP MODULE ADMINISTRATION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
ade ajibade
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key7404232
MDR Text Key104606942
Report Number9616066-2018-00398
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K944320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2432-0007
Device Catalogue Number2432-0007
Other Device ID Number10885403232329
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received04/09/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BAXTER 1000ML BAG, THERAPY DATE UNK
Patient Outcome(s) Required Intervention;
Patient Age14 YR
-
-