• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS INC. TRILOGY 100; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS INC. TRILOGY 100; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 1054096
Device Problems Premature Discharge of Battery (1057); Charging Problem (2892)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 03/15/2018
Event Type  malfunction  
Event Description
The manufacturer received information alleging a ventilator's batteries were depleting.There was no harm or injury reported.The device has not yet been returned to the manufacturer for evaluation.At this time, we are unable to confirm the alleged malfunction.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Manufacturer Narrative
It was initially reported a ventilator's batteries were depleting.There was also an allegation the device's settings changed on their own.The ventilator was returned to the manufacturer for evaluation and the customer's complaints could not be confirmed or duplicated.The device performed to manufacturer's specifications.The ventilator's downloaded event logs were reviewed by the manufacturer.The device maintains the event logs in non-volatile memory.An event, as logged by the device, includes every keypress, alarm or failure of the device to remain within specifications.There were no events logged related to device performance or device malfunctions.If a device malfunction occurred, the event logs would have contained entries related to the malfunction.The logs confirm that only manual settings changes had occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 100
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
adam price
312 alvin drive
new kensington, PA 15068
7243349303
MDR Report Key7406268
MDR Text Key104829376
Report Number2518422-2018-00864
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00606959022744
UDI-Public00606959022744
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K083526
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 06/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1054096
Device Catalogue Number1054096
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/29/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/05/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-