• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
A follow-up report will be submitted upon completion of the investigation.
 
Event Description
It was reported that the ventilator had a battery failure.It was reported that the event occurred during clinical use.There was no report of patient harm.
 
Manufacturer Narrative
The field service engineer (fse) evaluated the device and verified the reported condition.The device showed error.The fse replaced the battery to address the issue.The ventilator was not in use on a patient at the time of the event occurred.The ventilator passed all tests and operated within the manufacturing specifications.
 
Event Description
It was reported that the ventilator had a battery failure.The ventilator was not in use on a patient at the time of the event occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
robert corning
3000 minuteman road
andover, MA 01810
9786871501
MDR Report Key7410354
MDR Text Key104876808
Report Number2031642-2018-00780
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838025776
UDI-Public(01)00884838025776
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/19/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-