Manufacturer Narrative
Analysis summary: analysis confirmed the customer comment of a broken ventricular output connector but was unable to confirm the comment that the pacing cable was unable to be connected.It was additionally noted that the hanger was broken.The electrical and mechanical parts were inspected and all found defective parts were replaced and all other identified issues were resolved.The device was recalibrated and functionally tested and passed its final quality assurance tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the external pulse generator's connector for the ventricular cable was broken and that it was not possible to insert the pacing cable.The generator was returned for service.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7415874
MDR Text Key105311825
Report Number3004593495-2018-00219
Device Sequence Number1
Product Code DTE
UDI-Device Identifier00643169725126
UDI-Public00643169725126
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received02/01/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/05/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-