Model Number N/A |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/16/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that when the surgeon tried to push the button everything disassembled.Attempts have been made and additional information on the reported event is unavailable.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.
|
|
Manufacturer Narrative
|
Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the returned devices showed that the quick connect handle plunger was fractured and disassembled.The spring and fractured portion of the plunger are not returned.Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|