• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Device Problem Signal Artifact/Noise (1036)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Product analysis: at analysis the unit failed incoming testing on the serial number reading and the ventricular pace pulse noise tests and this was attributed to the main board being defective.The unit was cleaned and inspected, the main board was replaced and the unit was tested and calibrated on the automated test system and passed.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The external pulse generator was returned for preventive maintenance and subsequently tested out of specification during manufacturer's analysis.There was no patient involvement.
 
Manufacturer Narrative
Failure analysis was performed on the main board.Visual inspection revealed no anomalies.The main board was assembled into a golden unit.The board was then run on an automated test console.The board failed ventricular pulse noise test.The ventricular pulse noise was measured on the bench at 49.2 mv.The measured ventricular pulse noise was within the requirements of the ventricular pulse noise test (0-100 mv).The error logs were analyzed, no errors found.It was confirmed that the unit fails the ventricular pulse noise test during automated testing, but ventricular pulse noise level measured is within device specifications.Conclusion: no defect found.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7417323
MDR Text Key105264818
Report Number3004593495-2018-00252
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received06/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/13/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-