Product analysis: at analysis the unit failed incoming testing on the serial number reading and the ventricular pace pulse noise tests and this was attributed to the main board being defective.The unit was cleaned and inspected, the main board was replaced and the unit was tested and calibrated on the automated test system and passed.If information is provided in the future, a supplemental report will be issued.
|
Failure analysis was performed on the main board.Visual inspection revealed no anomalies.The main board was assembled into a golden unit.The board was then run on an automated test console.The board failed ventricular pulse noise test.The ventricular pulse noise was measured on the bench at 49.2 mv.The measured ventricular pulse noise was within the requirements of the ventricular pulse noise test (0-100 mv).The error logs were analyzed, no errors found.It was confirmed that the unit fails the ventricular pulse noise test during automated testing, but ventricular pulse noise level measured is within device specifications.Conclusion: no defect found.If information is provided in the future, a supplemental report will be issued.
|