• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V2R
Device Problem Material Protrusion/Extrusion (2979)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The scope was returned to olympus for evaluation. The evaluation confirmed the reported scope issue. Olympus performed a visual inspection and found a portion of the bending section with a sharp edge and exposed/protruding bending skeleton metal tab causing the scope to fail leak testing. Upon the removal of the bending section cover, the bending section skeleton was found completely broken/detached with a sharp edge. Based on similar reported events and investigation findings, the cause of the protruding / lifting skeleton metal tab could be attributed to the operator¿s technique. The original equipment manufacturer (oem) performed investigations related to this device issue. As a result, the oem has conducted a field corrective action including a distribution of ¿instructions for safe use¿ to mitigate the potential risk of patient injury. The ¿instructions for safe use¿ provides several warning statements in an effort to prevent equipment damage and patient injury. ¿if any of the following conditions occur during an examination, immediately stop the examination and withdraw the endoscope from the patient. If the up/down angulation control lever does not move. If the angulation control mechanism is not functioning properly. Visually inspect the bending section for no metallic parts protruding from the bending section. Visually inspect the bending section for bends, twist, or other irregularities while the bending section remains straight. Visually inspect the bending section for abnormal bending shape, or other irregularities. Continued use of the endoscope under these conditions could result in patient injury, bleeding, and/or perforation. ".
 
Event Description
Olympus was informed that during the reprocessing process of the scope, the bending section was found with a broken rib/skeleton protruding through the bending section rubber. There was no patient/user injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameURETERO-RENO VIDEOSCOPE
Type of DeviceURETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
connie tubera
2400 ringwood avenue
san jose, CA 95131
408935-512
MDR Report Key7425455
MDR Text Key105403206
Report Number2951238-2018-00237
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK072957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation
Type of Report Initial
Report Date 04/12/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberURF-V2R
Device Catalogue NumberURF-V2R
Device Lot NumberN/A
Other Device ID Number4953170343612
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2018
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/22/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1251-2017

-
-