BAXTER HEALTHCARE - ENGLEWOOD 500 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938738
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 02/22/2018
Event Type  malfunction  
Manufacturer Narrative
The device was returned and an evaluation is complete.Functional leak testing was performed by filling the bag with water and a leak was noted at the spike port cap.Visual inspection was performed and noted the fill port tubing was detached.The tubing was closed off and the bag was filled with tap water.There was a leak noted at the spike port cap.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The cause was not determined.This issue is being further investigated.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
During evaluation on a returned 500ml exactamix eva tpn bag it was found the device had leaked from the spike port cap.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
500 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE - ENGLEWOOD
englewood CO
Manufacturer (Section G)
BAXTER HEALTHCARE - ENGLEWOOD
14445 grasslands dr
englewood CO 80112
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7425680
MDR Text Key105305147
Report Number1416980-2018-02075
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477220
UDI-Public(01)00085412477220
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date05/31/2020
Device Catalogue NumberH938738
Device Lot Number60069109
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2018
Date Manufacturer Received03/20/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-