DEPUY SYNTHES PRODUCTS LLC ELECTRIC PEN DRIVE 60,000 RPM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 05.001.010 |
Device Problem
Device Inoperable (1663)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/20/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported from (b)(6) that during an unknown surgical procedure it was observed that the electric pen drive device was out of order.It was not reported if there was a delay in the procedure due to the event.It was reported that an unspecified spare device was available for use to complete the procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.The device was evaluated and it was determined that the control unit of the electric pen drive device was not functioning and was defective.It was noted that the pen drive was without function.It was further determined that the device failed pretest for check function and direction of rotation.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to improper maintenance, which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|