MAUDE Adverse Event Report: SMITH & NEPHEW, INC. CAMERA HEAD LENS INT SYS; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. CAMERA HEAD LENS INT SYS; ARTHROSCOPE Back to Search Results
Model Number 72203967
Device Problem Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/15/2018
Event Type  malfunction  
Event Description
It was reported that the buttons were stuck in the camera head.No patient injury or significant time delay reported.No back up device was available.
 
Manufacturer Narrative
The device was received for evaluation.There was a relationship found between the returned device and the reported incident.Visual inspection of the returned device did not identify any issues.Functional evaluation revealed that all 3 buttons were nonfunctional; there was no response when pressed.None of the buttons were found to be stuck.The reported malfunction could not be duplicated during the functional testing process.Further investigation was not possible as corrosion around the threads at the base of the housing prevented disassembly of the unit.Factors that could have contributed to the reported event include a variation in distances between components or an installation error resulting in inconsistent button activation.A corrective action has been initiated to mitigate future recurrence of similar events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAMERA HEAD LENS INT SYS
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7427163
MDR Text Key105417992
Report Number1643264-2018-00282
Device Sequence Number1
Product Code HRX
UDI-Device Identifier00885556629833
UDI-Public(01)00885556629833
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72203967
Device Catalogue Number72203967
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/03/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-