MERGE HEALTHCARE MERGE UNITY PACS PICTURE AND ARCHIVING COMMUNICATIONS SYSTEM Back to Search Results
Model Number UNITY PACS R10.0.436.77
Device Problems Computer Software Problem (1112); Loss of Data (2903)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/15/2018
Event Type  malfunction  
Manufacturer Narrative
Merge unity pacs is designed to display a message to the user when an exam report fails to upload to the image server. In this instance, there was no evidence that the physician received the message warning that the report did not upload. It was not possible to determine if the message displayed on the screen for the physician when the report was first read and saved. Unity tech investigation showed the database was updated but the report did not upload to the server; the report header, body, and report files were not present. The exam was re-read by a physician. Station logs had truncated, therefore were unavailable for review and further investigation. As a result, no other review/investigation can be performed for cause.
 
Event Description
Merge unity pacs is a medical image and information management system that is used for viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems. On march 15, 2018, the customer reported an exam report that was dictated on (b)(6) 2018 was not saved and is missing. Unity tech investigation showed the report did not upload successfully, and will have to be recreated. The audit trail does show a "report save" event but header, body, and report files were not present. Station logs had truncated, and were unable to be reviewed. The exam was read on (b)(6) 2018. While images are available, a finalized report not being available for subsequent review by the patient's physician could potentially result in a delay in care that could lead to harm. However, the customer did not allege any harm, serious injury or death as a result of this issue. Reference complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMERGE UNITY PACS
Type of DevicePICTURE AND ARCHIVING COMMUNICATIONS SYSTEM
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
meg mucha
900 walnut ridge drive
hartland, WI 53029
MDR Report Key7427805
MDR Text Key105697548
Report Number2183926-2018-00025
Device Sequence Number1
Product Code LLZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K041935
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation
Type of Report Initial
Report Date 03/15/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberUNITY PACS R10.0.436.77
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received03/15/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-