One jelco® viavalve® safety iv catheter was returned for analysis.
The customer's reported problem was "the ivc was not threading properly and valve leaking.
" the catheter sub-assembly was not included; so the valve could not be evaluated.
Blood was present in the guard.
It would appear that the excessive presence of blood may be a deployment training issue as the deployment chassis would have to be left attached to the catheter hub for a long period of time for the amount of blood to leak.
When manually advancing the assembly with slight downward pressure to simulate clinical use, it was observed that advancement was difficult and uneven in feel.
However, this may be due to the presence of blood residue.
Regarding the leakage, based on these observations and the lack of a beveled acam to examine, it is unconfirmed that the complaint was due to a manufacturing non-conformance.
Regarding the threading complaint, based on these observations the presence of it is unconfirmed that the complaint was due to a manufacturing non-conformance.
The manufacturing processes were reviewed and considered adequate and correct.
|