Catalog Number 2C7557 |
Device Problem
Cut In Material (2454)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/22/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device was received and evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the blue slide clamp of a vented paclitaxel set severed the tubing which resulted in a leak.The slide clamp severed the tubing during patient setup.The line had been primed and the patient was connected for therapy when the severed tubing was identified.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
The device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional, pressure and clear passage under water testing were performed and the device was found to be within the product specification range.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Search Alerts/Recalls
|