DEPUY SYNTHES PRODUCTS LLC RECON SAGITTAL SAW WITH KEY FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 05.001.240 |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/26/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|
Event Description
|
It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the saw attachment device blade holder was broken.It was reported that there were no delays in the surgical procedure.It was not reported whether a spare device was available for use.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the device coupling on the tool side was not functioning, defective, bearings, seal worn out and clamping head worn out.It was also noted that the device failed pre-test for check the saw blade coupling.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to improper maintenance, which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|